Report a non-conforming event
This page documents the Report Non-Conformity Event form introduced with the Quality Assurance menu in OpenELIS 3.2.3.0. If your site is on an earlier release, use PART 11: Non Conforming Events (page 278888457) instead.
How-to · Laboratory staff
At a glance
Purpose | Record a quality event, such as a damaged, mislabelled or delayed sample, so it can be investigated and followed through to closure. |
|---|---|
You'll end up with | An open NCE with its own number, listed under All NCEs and linked to the affected sample. |
Who does this | Any member of laboratory staff who notices the problem. Access is set by user permissions and can be configured by your administrator. |
Before you start | The lab number of the affected order. On this release an NCE is only saved when a sample is linked to it (see Tips). |
Where | Main menu → Quality Assurance → QMS & Improvement → Non-Conform → Report Non-Conforming Event, or Report NCE on the All NCEs page. |
The form has five numbered sections: Reporter & Event Context, Classification, Details, Attachments and Link to Samples / Results. Once submitted, the event appears under All NCEs, where it is acknowledged, assigned and investigated.
Step 1: Open the form and set the context
Open Main menu → Quality Assurance → QMS & Improvement → Non-Conform → Report Non-Conforming Event. Under 01 Reporter & Event Context the NCE Number is already filled in (for example NCE-2026-00012), Reporter Name shows your user name and Date of Event defaults to today. Change the date if the event happened earlier, and choose the Reporting Unit, the lab unit reporting the event.
Step 2: Classify the event
Choose a Category: General, Order, Sample, Analysis or Post-Analytical.
Choose a Subcategory. The list depends on the category. For Sample it offers, among others, Broken Tube/Container, Coagulated Sample, Hemolytic sample, Insufficient Amount, Mislabeled/unlabeled specimen, Specimen Receipt Delay and Wrong temperature.
Click the Severity card that fits: Critical (patient safety risk, regulatory violation), Major (significant quality or operational impact) or Minor (limited impact, easily corrected). The chosen card is highlighted.
Reporter and event context, with the Sample category, the Hemolytic sample subcategory and Minor severity chosen.
Step 3: Describe what happened
Under 03 Details, give the event a short Title and a Description (required). Record what was done straight away in Immediate Action Taken. Suspected Causes and Proposed Action are optional; the form notes that the suspected cause is an initial assessment only, with the formal root cause determined during the investigation, and that the proposed action is separate from the CAPA actions assigned later.
The Details section filled in for a haemolysed EDTA sample.
Section 04 Attachments accepts up to five files of up to 10 MB each, by drag and drop or by clicking the upload area.
Step 4: Link the affected sample
Under 05 Link to Samples / Results, choose Search By (Last Name, First Name, Patient Identification Code or Lab Number), type the value in Text Value and click Search.
In the results, tick the specimen under Specimen type, for example Whole Blood (DEV01260000000000494-1).
Click Link Selected Samples. The lab number appears as a tag at the top of the section; its x removes it.
The sample linked: its lab number shows as a tag above the search fields.
Step 5: Submit
Click Submit NCE. The message Save was successful appears and the form clears, ready for the next event. Cancel leaves the form without saving.
After Submit NCE: the success message, and the form cleared for the next event.
Step 6: Find the event under All NCEs
Open Main menu → Quality Assurance → QMS & Improvement → Non-Conform → All NCEs. The page lists all open NCEs across the laboratory, with counts for Critical, Major, Minor and Overdue. Narrow the list with the search box and the status (Open, Under Investigation, CAPA, Completed), category and severity filters, and, in 3.3.0, the source filter (All sources or EQA-triggered, for NCEs raised automatically from unacceptable EQA scores); Clear All resets them. Each row shows the NCE number, the linked sample, the status, the title, the category and subcategory, and how many days the event has been open.
All NCEs searched for the new event: NCE-2026-00012, Open, Sample - Hemolytic sample.
Click the row to open it. Event Details shows the description, the immediate action and the linked samples, with tabs for Investigation, CAPA and History, and the actions Acknowledge, Assign To and Add Note.
The new NCE opened, showing Event Details and the Acknowledge, Assign To and Add Note actions.
Field reference
Field | Notes |
|---|---|
NCE Number | Assigned by OpenELIS and shown before you submit. |
Reporter Name | Your user name. |
Date of Event | Required. Defaults to today. |
Reporting Unit | Required. The lab unit reporting the event. |
Category | Required. General, Order, Sample, Analysis or Post-Analytical. |
Subcategory | The list depends on the category. |
Severity | Required. Critical, Major or Minor. |
Title | A short description, used as the heading in All NCEs. |
Description | Required. What happened. |
Immediate Action Taken | What was done straight away. |
Suspected Causes | Optional initial assessment. |
Proposed Action | Optional; separate from CAPA actions. |
Attachments | Optional. Up to five files, 10 MB each. |
Link to Samples / Results | Search by patient name, patient identification code or lab number, tick the specimen and click Link Selected Samples. |
Tips
Known defect (OGC-1379): section 05 is marked Optional, but an NCE with no linked sample is not saved. Submit NCE shows Save failed and nothing is recorded. Link the affected sample before submitting.
The All NCEs severity filter also offers Low, which the report form does not.
The older screens View New Non-Conforming Events and Corrective actions are still in the Non-Conform menu. This page does not cover them.
Not yet verified
Uploading attachments.
Acknowledge, Assign To and Add Note, and the Investigation, CAPA and History tabs.
Linking by patient name or identification code, and linking more than one sample.
Provenance
Verified by Claude (Cowork) for Casey Iiams-Hauser · Date 2026-10-09 · Instance testing · OpenELIS version 3.2.3.0 (authored and driven 2026-09-28; re-driven end-to-end 2026-10-09 on the 3.3.0-SNAPSHOT build now on testing) · Review OGC-1381
Gates applied: ☑ workflow sentence ☑ first real step ☑ step↔screenshot fidelity ☑ driven live end-to-end ☑ process gates
Instance | testing.openelis-global.org, OpenELIS 3.2.3.0 |
|---|---|
Date | 28 September 2026; review re-run 9 October 2026 |
Driven | Filled and submitted the form twice with a linked sample (NCE saved, NCE-2026-00012 opened from All NCEs), and twice without one (Save failed, API error labOrderNumber is required). Review on 9 October: submitted without a sample (Save failed, OGC-1379 still present), then linked a sample by lab number and submitted (Save was successful, form cleared); the event appeared Open under All NCEs and opened to Event Details with the linked sample. |
Observed only | Attachments area, the All NCEs filters and counts, the Investigation, CAPA and History tabs and the detail actions. |
Jira | QA Menu Release OGC-682; defect OGC-1379; review OGC-1381 |
Replaces | PART 11: Non Conforming Events (278888457), now under LEGACY Versions for sites on earlier releases. |
Drift contract | nce-report |
Updated 13 October 2026: the All NCEs source filter (All sources, EQA-triggered) noted from testing.openelis-global.org, 3.3.0-SNAPSHOT. OGC-1379 is still in Backlog, so the Tip about saving without a linked sample stands.