Statistical QC: overview and daily review

Statistical QC: overview and daily review

Work in progress. This page is published before its review is finished so you can use it now. It is being reviewed under OGC-1359, so details may still change. If something on your screen does not match, add a comment to this page.

Concept and how-to · Quality officers, supervisors and bench staff

Status: draft for review. Written against pngtest.openelis-global.org, version 3.2.2.0, 27 September 2026; dashboard re-checked on testing.openelis-global.org, 3.3.0-SNAPSHOT, 12 October 2026. Quality control is being reworked around Runs (OGC-1200), so expect this page to change. See Provenance.

At a glance

Purpose

Know where statistical quality control lives in OpenELIS and check each day whether any control has failed.

You'll end up with

A view of instrument QC status, bench and rapid-test controls, and any violations that need action.

Who does this

Quality officers and supervisors, and bench staff who run controls. Access is set by user permissions and can be configured by your administrator.

Before you start

At least one active control lot (see Receive and activate a control lot).

Where

Main menu → Quality Assurance → Statistical QC.

What is in Statistical QC

Menu item

What it is for

Menu item

What it is for

QC Dashboard

Status summary with three tabs: Instruments, Alerts and Bench QC.

QC Alerts

Opens the dashboard on its Alerts tab.

QC Lot Management

Control lots and their target statistics.

Rule Configuration

The Westgard rule sets configured per analyzer and test ("Configured Rule Sets").

Reagent QC

Not yet available. The page says that it is a future feature: the design exists, but the build is not yet scheduled.

Analyzer Manual QC

Not yet available. Same future-feature notice.

Figure 1. Analyzer Manual QC shows a future-feature notice. Reagent QC shows the same kind of notice.

Figure 1. Analyzer Manual QC shows a future-feature notice. Reagent QC shows the same kind of notice.

Check QC status each day

  1. Open QC Dashboard. The four tiles at the top show Instruments In Control, Warning Status, Out of Control and % Instruments In Control (called Overall QC Pass Rate on 3.2.2.0). Last Updated shows when the figures were read; Refresh and Export sit beside it. Export opens Export QC Data, where you choose the Instrument, the Control Level (All levels by default) and a Start Date and End Date, then Export CSV or Export PDF. On the test server, with no instrument to choose, both export buttons were unavailable.

  2. The Instruments tab lists each instrument with its Type, Location, Status, Analytes, Recent Violations, Last Update and Actions. Use Search instruments to find one.

Figure 2. The QC Dashboard on its Instruments tab (3.3.0). This instance has no instruments connected, so the list is empty.

Figure 2. The QC Dashboard on its Instruments tab (3.3.0). This instance has no instruments connected, so the list is empty.

  1. Open the Bench QC tab to see controls recorded without an analyzer, such as rapid-test controls. Each row shows the Lab unit, Test, Source, Controls run, how many Failed, and when the last one was run. Filter with Source.

Figure 3. The Bench QC tab with one rapid-test control run for HIV Rapid Test in Serology, marked as failed.

Figure 3. The Bench QC tab with one rapid-test control run for HIV Rapid Test in Serology, marked as failed.

  1. Open the Alerts tab (or QC Alerts in the menu). Active QC Violations lists violations not yet acknowledged; Acknowledged Violations lists those already dealt with. Choose the Time Period to look back over, for example Last 72 Hours.

Figure 4. The Alerts tab with no active or acknowledged violations in the last 72 hours.

Figure 4. The Alerts tab with no active or acknowledged violations in the last 72 hours.

Tips

  • A failed bench control does not appear on the Alerts tab: with the failed HIV Rapid Test control in Figure 3 on the dashboard, Alerts showed no violations. Check the Bench QC tab as well as Alerts.

  • These pages show Analyzers › Quality Control Dashboard in the breadcrumb even when reached from Quality Assurance. This is a known inconsistency, not a different page.

  • How rapid-test and manual control results are recorded is not yet documented in this manual.

Not yet available

  • Reagent QC and Analyzer Manual QC (future-feature notices in the menu).

  • Runs: one unit of QC, reagent lot and result review across analyzer, workplan and manual results (OGC-1200, planned).

Not yet documented

  • Recording a rapid-test or manual control result.

  • Reading an instrument's Levey-Jennings chart and acknowledging a violation (needs a connected analyzer with QC data).

  • Editing a Westgard rule set.

Provenance

Instance

pngtest.openelis-global.org, version 3.2.2.0

Date

27 September 2026

Observed

Menu, dashboard tiles and tabs, Bench QC row (recorded by another user earlier the same day), Alerts tab, Rule Configuration list, the two future-feature pages.

Driven

Tab switching; created control lot 4521A-L1 and confirmed it did not add a rule set.

Not verified

Instrument rows, violation acknowledgement, Export contents, rule editing.

Feature

QA Menu umbrella OGC-682 (Feature Doc OGC-1213), Statistical QC rehome OGC-689, manual and RDT QC OGC-1147

Updated

12 October 2026 on testing.openelis-global.org, 3.3.0-SNAPSHOT: menu items and routes unchanged (QC Dashboard, QC Alerts, QC Lot Management, Rule Configuration, Reagent QC and Analyzer Manual QC still future features); the fourth tile is now % Instruments In Control; Export opens Export QC Data (observed, not exported: no instruments); Bench QC empty on testing (Figure 3 is from pngtest in September). Figure 2 recaptured.

Reviewer

Not yet reviewed