General Configurations reference

General Configurations reference

Reference · Administrators (technical)

Part of: Test Catalog Management: workflow overview

Reviewed (OGC-1299, 28 September 2026). Captured from a live instance on 15 September 2026. Updated 12 October 2026 against 3.3.0-SNAPSHOT: settings added in 3.3.0 are marked New in 3.3.0, the one removed is marked 3.2.x only, and reference values that changed show both.

At a glance

Purpose

Look up what each General Configuration setting controls, and what it is set to.

You'll end up with

An understanding of which behaviours at your site are configuration-driven.

Who does this

Administrators, usually with technical support. Access is set by user permissions and can be configured by your administrator.

Before you start

Nothing. Read before changing anything: several of these settings change what other staff see on screen.

Where

Main menu → Admin → General Configurations

Read this first

These settings change what other pages in this manual describe. A how-to page shows one configuration's version of a screen. If a field described in a task page is missing at your site, or a field appears that no page mentions, the cause is usually a setting on this page rather than an error.

The settings most likely to cause that confusion are flagged Changes the UI below.

Every screen here works the same way: select a row, click Modify, change the value, save. Values are shown as they were found on the reference instance; yours will differ.

Order Entry Configuration

  1. Click Admin

Admin August.png
Admin
  1. Click General Configurations.

General Conf.png
General configurations

3. Click Order Entry configuration.

order entry conf.png
Order entry configuration
  1. Select a row, then Modify an Order Entry Configuration.

order entry config.png
Modify order entry

Setting

What it controls

Reference value

Setting

What it controls

Reference value

auto-fill collection date/time

Whether specimen collection date and time are pre-filled

false

billingRefNumber

Whether a billing reference number is tracked

false

billingRefNumberLocalization

The label used for the billing reference field, per language

English: URAP Number / French: N° URAP

contactTracingEnabled

Changes the UI. Shows contact-tracing fields

false

eqaEnabled

Changes the UI. Shows the EQA checkbox, letting a sample be marked as an EQA sample

true

external orders

Allows external sites to send electronic orders

false

gpsCoordinatesEnabled

Changes the UI. Shows GPS coordinate fields for the collection location

false

gpsRequiredAccuracyMeters

Maximum acceptable GPS accuracy in metres; 0 disables the check

100

gpsTimeoutSeconds

Timeout for GPS location requests

10

Patient age required

New in 3.3.0. Changes the UI. Makes the patient date of birth required when a patient is created or edited

true (3.3.0)

Patient sex required

New in 3.3.0. Changes the UI. Makes the patient sex required when a patient is created or edited

true (3.3.0)

Program

Changes the UI. Whether a program is part of an order

true

requesterRequired

Makes Search Requester and Requester name required rather than optional

false

restrictFreeTextProviderEntry

Prevents entering new providers during sample entry

false

restrictFreeTextRefSiteEntry

Prevents entering new referring sites during sample entry

false

restrictFreeTextRequestorEntry

Prevents entering new Environmental/Vector requestor contacts

false

restrictFreeTextSampSiteEntry

Prevents entering new Environmental/Vector sampling sites

false

sampleAcceptCheck.clinical

Enforcement level for the Clinical sample acceptance checklist: MANDATORY, OPTIONAL or OFF

OPTIONAL

sampleAcceptCheck.environmental

Same, for Environmental

OPTIONAL

sampleAcceptCheck.vector

Same, for Vector

OPTIONAL

showFaxFields

New in 3.3.0. Changes the UI. Shows fax fields for facilities and providers in order entry

false (3.3.0)

statTurnaroundMinutes

New in 3.3.0. Minutes after receipt a STAT order may go without a result before an Overdue STAT alert is raised

60 (3.3.0)

trackPayment

Changes the UI. Tracks patient payment for services

true

validateAccessionNumber

Validates the format of lab numbers

true

The four restrictFreeText… settings are the reason a site may be unable to type a new provider, referring site or sampling site during order entry and must pick from the list instead.

Result Entry Configuration

Admin → General Configurations → Result Entry Configuration

result entry conf.png
Result Entry Configuration
04-cfg-result-entry (1).png
Result entry

Setting

What it controls

Reference value

Setting

What it controls

Reference value

resultsEntryUnifiedRoute

3.2.x only; removed in 3.3.0. Changes the UI, substantially. When true, the unified Results Entry worklist replaces the legacy result entry pages and legacy routes redirect

true

alertWhenInvalidResult

Alerts when a result is entered outside the valid range

false

allowBulkReleaseClear

Lets validators release all Clear-lane results at once under one e-signature, rather than one at a time

true

allowResultRejection

Changes the UI. Lets a technician reject an individual test and pick a reason

false

autoFillTechNameBox

Adds an autofill box for the technician name

false

autoFillTechNameUser

Autofills the technician name with the logged-in user

false

customCriticalMessage

The critical result message text

Set new critical result message

modify results note required

Requires a note when results are modified

false

modify results role

Requires a separate role to modify results

false

pathology.stage.<STAGE>.enabled

New in 3.3.0. Seven settings, one per pathology stage (COVERSLIPPING, DECALCIFICATION, EMBEDDING, MICROTOMY, PROCESSING, STAINING, UNDER_REVIEW). If true, cases pass through that stage; if false, it cannot be chosen and is not applicable

true (all seven)

qcFailBlocksValidation

New in 3.3.0. Blocks release of results covered by an open QC failure

false (3.3.0)

restrictFreeTextMethodEntry

Prevents entering new methods during result entry

false

ResultTechnicianName

Changes the UI. Requires the technician name when entering results

false

retestNoteRequired

Requires a validator to write a note when sending a result back for retest

false

roleForPatientOnResults

Restricts patient information to the correct role

false

showValidationFailureIcon

Shows an icon on the results page when an analysis fails validation

true

validate all results

Requires all results to be validated, including normal ones

true

validateTechnicalRejection

Makes technician-rejected results ready for validation

true

resultsEntryUnifiedRoute deserves particular attention on 3.2.x. There it decides which Results Entry screen your staff see, and it is a configuration toggle rather than a version difference: two sites on the same release can differ. Check it before following any Results Entry page. In 3.3.0 the setting is gone: every site uses the single Results worklist, and the old result entry addresses open it.

Patient Entry Configuration

Admin → General Configurations → Patient Entry Configuration

patient entry conf.png
Patient Entry Configuration.
patient entry config.png
Patient entry configuration

 

Setting

What it controls

Reference value

Setting

What it controls

Reference value

Allow duplicate national ids

Whether two patients may share a national ID

true

Allow duplicate subject number

Whether two patients may share a subject number

true

default nationality

Nationality pre-selected for new patients; must match a value in the nationality list

(empty)

National ID required

Changes the UI. Makes national ID required

true (3.2.2.0); false (3.3.0)

patientGpsCaptureEnabled

Changes the UI. Enables GPS capture on patient registration

false

Patient ID required

Makes a patient ID required

false

Subject number required

Makes a subject number required

false

supportPatientNationality

Changes the UI. Whether patient nationality is captured

true

useNewAddressHierarchy

Uses the configurable address hierarchy instead of the legacy Health Region / District

true

useNewAddressHierarchy changes the address fields on patient registration, so a page describing one hierarchy will not match a site using the other.

Validation Configuration

Admin → General Configurations → Validation Configuration

This is not about result validation. Despite the name, it sets which characters are accepted in name and identifier fields. Result validation settings live under Result Entry Configuration above.

validation entry conf.png
Validation configuration

 

06-cfg-validation.png
Validation Configuration: the allowed character sets.

Setting

What it controls

Reference value

Setting

What it controls

Reference value

firstNameCharset

Characters allowed in first name fields

.'a-zàâçéèêëîïôûùüÿñæœ -

lastNameCharset

Characters allowed in last name fields

.'a-zàâçéèêëîïôûùüÿñæœ -

patientIdCharset

Characters allowed in patient identifier fields

a-z0-9/àâçéèêëîïôûùüÿñæœ

userNameCharset

Characters allowed in username fields

a-zàâçéèêëîïôûùüÿñæœ ._@-

If a patient name is rejected on entry for no obvious reason, these are why. Note that the reference values include accented Latin characters but no non-Latin scripts.

NonConformity Configuration

Admin → General Configurations → NonConformity Configuration

non conformity conf.png
Non conformity configuration
07-cfg-nonconformity.png
NonConformity Configuration.

Setting

What it controls

Reference value

Setting

What it controls

Reference value

Collection as unit

Whether sample collection can be reported as the place a non-conformity occurred

true

Reception as unit

Whether reception can be reported as that place

true

sample id required

Whether a requester sample ID is required

true

sortQaEvents

Sorts QA events in lists

false

WorkPlan Configuration

Admin → General Configurations → WorkPlan Configuration

workplan config.png
Workplan configuration
08-cfg-workplan.png
WorkPlan Configuration.

Setting

What it controls

Reference value

Setting

What it controls

Reference value

next visit on workplan

Changes the UI. Shows the next visit date on the workplan

false

results on workplan

Changes the UI. Leaves space for results on the workplan

false

subject on workplan

Changes the UI. Uses the subject number on the workplan

false

All three change what a printed workplan contains, so a workplan at your site may look unlike the examples elsewhere in this manual.

Printed Report Configuration

Admin → General Configurations → Printed Report Configuration

printed report confg.png
Printed report configuration
09-cfg-printed-report.png
Printed Report Configuration: report identity and header images.

These set the identity and images used in report headers and footers. All were empty on the reference instance except reportPageNumbers and, on 3.3.0, reportPaperSize.

Setting

What it controls

Setting

What it controls

reportPaperSize

New in 3.3.0. Paper size of printed reports

A4 (3.3.0)

SiteName

Site name used on reports and in the header

additional site info

Extra text for the report header

lab director

The lab director's name

labDirectorName

Image for the lab director name

labDirectorTitle

Image for the lab director title

labDirectorSignature

Image for the lab director signature

headerLeftImage

Image for the left of the report header

headerRightImage

Image for the right of the report header

reportPageNumbers

Whether reports carry page numbers (false on the reference instance)

Site Branding

Admin → General Configurations → Site Branding

Not a property table but a form. "Customize your organization's visual branding including logos and colors."

site branding conf.png
Site branding

Control

Notes

Control

Notes

Header Logo

Shown top-left in the application header. PNG, SVG or JPG, max 2 MB.

Login Page Logo

Shown on the login page. Has a Use same logo as header option.

Favicon

Shown in browser tabs and bookmarks.

Header Color

Background of the header bar. Accepts any CSS colour: hex, name or rgb().

Primary Color

Primary buttons, active menu items, links.

Secondary Color

Secondary buttons, borders, inactive states.

Reset to Default Branding is a destructive action and is marked as such on screen.

MenuStatement Configuration

Admin → General Configurations → MenuStatement Configuration

menu statement conf.png
Menu statement configuration

Empty on the reference instance: the table showed 0 items. The screen exists and what it is for has not been established.

menu statement config.png
Menu statement configuration

Site Information

Covered separately: it sits in this menu but has its own page, Configure site information.

Not yet documented

The following Admin entries also have no page: External Connections, Notification Trigger Configuration, FHIR Data Export Status, Notify User, Batch test reassignment and cancelation, Microbiology reference, Legacy Admin, and the Localization subtree.

Provenance

Status

Verified; promoted to the Administrator Guide on 13 October 2026

Instance

testing.openelis-global.org, OpenELIS 3.2.2.0

Captured

15 September 2026

Note

Values are from the reference instance and are examples, not recommendations.

Figures

7 attached, 1440×900 at 2x. Site Branding and MenuStatement Configuration were not captured at first, because the Site Branding route could not be determined without driving the menu; figures for both were added later.

Updated

12 October 2026 against testing.openelis-global.org, OpenELIS 3.3.0-SNAPSHOT: every General Configurations table read again. Added the settings new in 3.3.0 (Order Entry: Patient age required, Patient sex required, showFaxFields, statTurnaroundMinutes; Result Entry: seven pathology stage settings, qcFailBlocksValidation; Printed Report: reportPaperSize), marked resultsEntryUnifiedRoute as removed, and recorded that National ID required is false on the 3.3.0 test server. Other tables unchanged. Figures not recaptured.

Reviewer