Sample Acceptance Checklist
Part of: Test Catalog Management — workflow overview
How-to · Lab administrators (configuration) + order-entry staff (use)
The Sample Acceptance Checklist lets a lab define acceptance criteria (e.g. "Specimen container intact and properly labeled") that staff confirm on the QA Review step when entering an order. The checklist is configurable per domain (Clinical / Environmental / Vector) with a lab-wide fallback, and each domain chooses how strictly it is enforced: Mandatory makes recording an answer to each item a gate on accepting the sample, Optional records answers without blocking, and Off hides the checklist.
There are two sides: administrators configure the checklist (Admin → Order Entry Configuration → Sample Acceptance Checklist), and order-entry staff answer it on the QA Review step of the domain-specific Add Order flow (e.g. Add Clinical Order → Enter Order → Collect Sample → Label & Store → QA Review).
Note: this is distinct from the separate Reject Sample action on a sample (which records a rejection reason for resampling) — the Sample Acceptance Checklist is the configurable QA-Review checklist described here.
Part 1 — Configure the checklist (administrator)
Purpose | Define the acceptance criteria staff confirm, per domain, and how strictly they are enforced. |
|---|---|
You'll end up with | An active checklist (lab-wide and/or per-domain) and an enforcement level for each domain. |
Who does this | Lab administrators. Access is set by user permissions and can be configured by your administrator. |
Before you start | The acceptance criteria your lab wants to enforce. |
Where | Main menu → Admin → Order Entry Configuration → Sample Acceptance Checklist. |
Open Sample Acceptance Checklist. The submenu offers All domains (lab-wide) plus Clinical, Environmental, and Vector.
On All domains, add the criteria that apply everywhere: click Add item,
Type a Label, and Save. Each item shows with Order (reorder with the up/down arrows), Label, Active, and Edit.
a. To tailor a domain, open it (e.g. Clinical). It shows a Checklist enforcement dropdown — Mandatory (must answer to accept the sample), Optional (record but don't block), or Off (hide) — plus domain-specific items and the inherited lab-wide items (shown as "Active (inherited)").
b. Enforcement is set per domain, not per lab unit.
Per-domain enforcement dropdown and inherited lab-wide items (Clinical).
Part 2 — Answer the checklist on QA Review (Step 4)
Purpose | Complete each acceptance criterion for a sample on QA Review and accept it — or report a non-conformance / resample. |
|---|---|
You'll end up with | Each sample marked Accepted (or NCE / Resample) with its checklist answers recorded. |
Who does this | Order-entry / reception staff. Access is set by user permissions. |
Before you start | A configured checklist (Part 1) and an order in progress. |
Where | Order → Add Order → Add Clinical Order → Enter Order → Collect Sample → Label & Store → QA Review. |
Work an order through Enter Order → Collect Sample → Label & Store → QA Review (or jump to QA Review from the menu for an in-progress order).
On QA Review, the Intake Acceptance panel lists each sample (Lab #, Sample type, Transit, Eligibility). Select a sample to show its checklist; the active list matches the order's domain (e.g. "Active list: Clinical").
Each configured criterion appears with Pass / Fail / N/A. Answer each one.
Accept the sample. Under Mandatory enforcement the Accept sample button stays disabled until every item is answered (the gate); once answered, the sample flips to Accepted and Accept sample enables. Under Optional, answers are recorded but Accept isn't blocked; under Off, no checklist appears.
If a criterion fails, use Report NCE (log a non-conformance) or Resample instead of accepting.
The QA Review Intake Acceptance checklist: the configured item shown as Pass / Fail / N/A, with Accept sample disabled until answered.
Field reference
Field | Notes |
|---|---|
Checklist enforcement (per domain) | Mandatory — must answer every item to accept the sample (gate); Optional — records answers, doesn't block; Off — hides the checklist. |
Item — Label | The acceptance criterion text shown on QA Review. |
Item — Active | Whether the item is in force; lab-wide items show as "Active (inherited)" on a domain. |
Item — Order | Display order; reorder with the up/down arrows. |
Pass / Fail / N/A | The answer recorded per criterion on QA Review. |
Accept sample / Report NCE / Resample | Accept the sample (enabled once answered under Mandatory), log a non-conformance, or send for resampling. |
Notes & tips
Lab-wide first, then specialize: define common criteria on All domains; add domain-specific items only where a domain differs. Domains inherit the lab-wide list automatically.
Enforcement is per domain — changing Clinical does not change Environmental or Vector.
Turning a domain to Off hides the checklist for that domain's orders without deleting the configured items.
This is not the per-sample Reject Sample action (a separate control with its own rejection reasons).