PART 7 B: ANALYZER MANAGEMENT
ANALYZER MANAGEMENT
This guide covers the Analyzer Management module in OpenELIS Global v3.2.1.3. It explains how to configure laboratory analyzers, set up field mappings, monitor errors, and define analyzer types. The module is accessible from the Analyzers menu in the left navigation sidebar.
Navigation
The Analyzers menu contains three sub-pages:
Sub-page | URL | Purpose |
|---|---|---|
Analyzers List |
| Add, view, edit, and configure analyzer connections and field mappings. |
Error Dashboard |
| View and acknowledge errors generated by analyzer communication. |
Analyzer Types |
| Define and manage analyzer type definitions used when adding new analyzer instances. |
1. Analyzers List
The Analyzers List is the primary management screen. It shows all configured analyzers with their current status and provides controls to add, edit, and map new devices.
Dashboard KPIs
KPI | Description |
|---|---|
Total Analyzers | All analyzer records in the system, regardless of status. |
Active | Analyzers currently running and accepting incoming result messages. |
Inactive | Analyzers that have been manually set to inactive or are not yet activated. |
Plugin Warnings | Analyzers where the assigned plugin JAR is not currently loaded in the running system. Requires administrator attention. |
Analyzer Status Lifecycle
Each analyzer progresses through a lifecycle. Statuses marked as manually settable can be forced via the Status field in Add/Edit; others are assigned automatically by the system.
Status | Meaning | Manually Settable |
|---|---|---|
Inactive | Analyzer is disabled and will not process messages. | Yes |
Setup | Analyzer has been created and is being configured. Default status for new analyzers. | Yes |
Validation | Mappings have been saved and are being tested before going live. | Yes |
Active | Analyzer is online and processing messages. Set automatically when mappings are activated. | No (automatic) |
Error Pending | A communication or mapping error was detected. Check the Error Dashboard. | No (automatic) |
Offline | The analyzer is unreachable on the network. Connectivity will be re-tested automatically. No manual recovery action is available in the UI — the system retries the connection automatically. | No (automatic) |
Analyzer Table Columns
Column | Description |
|---|---|
Name | Display name assigned to the analyzer instance. A Plugin Missing badge appears if the assigned plugin JAR is not loaded. |
Type | The Analyzer Type assigned, which determines the plugin and protocol used. |
Connection | IP address and port number of the analyzer (format: |
Test Units | Lab units to which this analyzer's results will be routed (shown as "X unit(s)"). |
Status | Current lifecycle status (see Status Lifecycle above). |
Last Modified | Date and time of the most recent change to this analyzer's configuration. |
Actions | Per-row overflow menu (⋯). See Per-Analyzer Actions below. |
Adding a New Analyzer
Click the Add Analyzer button (top-right of the Analyzers List).
The Add New Analyzer form opens.
Fill in the required fields (see table below).
Click Test Connection to verify the IP and port are reachable before saving.
Click Save to create the analyzer, or Cancel to discard.
Field | Description |
|---|---|
Analyzer Name | A descriptive name for this analyzer instance (e.g., Hematology Analyzer 1). Required. |
Status | Manual lifecycle override. Options: Inactive, Setup (default), Validation. The status note reads: "Status transitions automatically based on analyzer state. Manual override to INACTIVE, SETUP, or VALIDATION only." |
Plugin Type | The software plugin that processes incoming messages. Options are loaded dynamically from installed analyzer plugins on the server. |
Analyzer Type | The analyzer type definition (see Section 4). Options are loaded from configured active analyzer types. Required. |
Protocol Version | Communication protocol version. Options: ASTM LIS2-A2 (default), HL7 v2.3.1, HL7 v2.5. |
IP Address | Network IP address of the analyzer (e.g., 192.168.1.10). |
Port Number | TCP port the analyzer listens on (default: 5000). |
Per-Analyzer Actions
Click the overflow menu (⋯) in the Actions column to access the following:
Action | Description |
|---|---|
Field Mappings | Opens the Field Mapping editor to configure how analyzer output fields map to OpenELIS test results (see Section 2). |
Test Connection | Attempts a live connection to the analyzer's IP and port and reports success or failure. |
Configure File Import | Opens the File Import Configuration form for this analyzer. Available for all analyzers regardless of protocol. |
Copy Mappings | Copies the complete field mapping configuration from another analyzer of the same type. |
Edit | Opens the analyzer configuration form to update connection settings, status, or type assignment. |
Delete | Permanently removes the analyzer and all its field mappings. This action cannot be undone. |
Configure File Import
The Configure File Import form sets up directory-based file polling for analyzers that deliver results as flat files (CSV, TSV, or Excel). It is available for all analyzers via Actions → Configure File Import.
Field | Description |
|---|---|
Analyzer | The analyzer this file import configuration applies to. |
File Format | Format of the result files: CSV (default), TSV, or Excel. |
Import Directory | Directory path the system will watch for new result files. |
File Pattern | Glob pattern to match incoming files (e.g., |
Archive Directory | Optional. Directory where successfully processed files are moved after import. |
Error Directory | Optional. Directory where files that failed processing are moved. |
Column Mappings | JSON object mapping file column names to OpenELIS field names (e.g., |
Delimiter | Character used to separate columns. Applies to CSV and TSV formats (default: comma). |
Has Header Row | When checked (default), the first row of the file is treated as column headers. |
Active | When checked (default), the import configuration is active and polling will occur. |
2. Field Mappings
Field Mappings define how data fields in analyzer messages are translated into OpenELIS test results. Mapping must be completed before an analyzer can go live. Access Field Mappings via Actions → Field Mappings on any analyzer row.
Field Mappings Page Overview
The Field Mappings page (/analyzers/{id}/mappings) is divided into a left panel (field list) and a right panel (mapping editor), with a set of summary KPIs and action buttons above.
Page KPIs:
KPI | Description |
|---|---|
Total Mappings | All field mappings configured for this analyzer. |
Required Mappings | Fields that must be mapped before the analyzer can be activated. |
Unmapped Fields | Fields received from the analyzer that have no mapping configured. |
Pending Codes | Analyzer test codes received that do not match any configured mapping. |
Plugin Configuration Snapshot: Shows the current plugin configuration as a JSON object. Useful for verifying that the plugin is reading the correct field definitions.
Pending Unmapped Codes: A list of analyzer test codes received from the physical analyzer that could not be matched to any mapping. Use this section to identify codes that need new mappings.
Left Panel — Analyzer Fields
Column | Description |
|---|---|
Field Name | The field identifier as received from the analyzer. |
ASTM Ref | The ASTM segment reference for this field (e.g., R.3.1). |
Type | Field type: Numeric, Qualitative, Text, Date/Time, Control Test, Melting Point, or Custom. |
Unit | Unit of measure for Numeric fields. |
Action | Per-field actions (appears when a field is selected). |
A status line below the table shows: X fields available • Mapped: X • Unmapped: X • Draft: X • Active: X.
The filter dropdown allows filtering by: All Fields, Mapped Only, Unmapped Only, Draft Only, Active Only.
Fields are loaded by the system when the analyzer sends messages. If no analyzer data has been received, the field list will be empty. Use Query Analyzer to request the field list directly from the device.
Right Panel — Mapping Editor
When a field is selected in the left panel, the right panel shows the mapping editor for that field. When no field is selected, it displays: "Select a field from the left panel to view or create mappings."
Analyzer Field Types
Field Type | Description |
|---|---|
Numeric | Quantitative result (e.g., hemoglobin). Supports unit mapping with optional conversion factor. |
Qualitative | Coded result (e.g., Positive/Negative). Requires qualitative code mapping to OpenELIS dictionary values. |
Text | Free-text result stored as-is. |
Date/Time | Timestamp field from the analyzer message. |
Control Test | Quality control field. Links to QC records rather than patient results. See QC Mapping below. |
Melting Point | Specific to molecular analyzers that report melting curve temperatures. |
Custom | User-defined field type for non-standard analyzer outputs. |
Saving and Activating Mappings
Action | Effect |
|---|---|
Save as Draft | Stores the mapping in a pending state. Does not affect live message processing. |
Save and Activate | Applies the mapping to live message processing immediately. Requires confirmation. Applies to new messages only — does not reprocess previous results. |
Retire Mapping | Marks an existing mapping as retired. Retired mappings no longer process new messages but are preserved for audit history. |
Important: When activating on an Active analyzer, a warning is shown and you must check a confirmation checkbox before proceeding.
Unit Mapping (Numeric Fields)
For Numeric fields, a Unit Mapping sub-form lets you map the analyzer unit to the OpenELIS unit with an optional conversion factor.
Field | Description |
|---|---|
Analyzer Unit | The unit string as it appears in the analyzer message (e.g., g/dL). |
OpenELIS Unit | The corresponding unit in OpenELIS, selected from the configured unit dictionary. |
Conversion Factor | A multiplier applied to the analyzer value to convert it to the OpenELIS unit (e.g., 1.0 for no conversion). |
Reject If Mismatch | When enabled, results whose reported unit does not match the mapped unit are rejected rather than imported. |
Qualitative Code Mapping (Qualitative Fields)
For Qualitative fields, each distinct coded value from the analyzer must be mapped to an OpenELIS dictionary code. One mapping must be designated as the default, which is used when an incoming value has no explicit match.
Copy Mappings
The Copy Mappings action duplicates all field mappings from another analyzer of the same type. A confirmation dialog is shown before copying. If the source analyzer has no mappings, the action is blocked with an error message.
Query Analyzer
The Query Analyzer button opens a panel that requests the field list directly from the physical analyzer over the network connection. This populates the left panel with available fields without waiting for the analyzer to send a result message.
The Query Analyzer panel shows:
State: Current status of the query job (pending, in_progress, complete, failed).
Progress: Visual progress indicator.
Connection Logs: Timestamped log entries (e.g.,
[08:20:04] Starting query job,[08:20:04] Connecting to analyzer at 192.168.1.100:5000).
Testing Mappings with a Raw Message
The Test Mapping button opens the Test Field Mappings panel. Use this to validate that your mapping configuration correctly parses real analyzer output before activating.
Element | Description |
|---|---|
Analyzer Information | Shows Analyzer Name, Analyzer Type, and current count of active mappings. |
Sample ASTM Message | Required text area for pasting a raw analyzer message. Maximum 10,240 characters. |
Show detailed parsing steps | When checked, the preview output includes step-by-step parsing details. |
Validate all mappings | When checked, all configured mappings are validated against the message, not just the parsed fields. |
Preview Mapping | Submits the message and displays parsed results. |
The results panel shows each parsed field with its raw value and how it maps to OpenELIS. If no message has been entered yet, it displays: "No preview results yet. Enter a sample ASTM message and click Preview Mapping."
3. Error Dashboard
The Error Dashboard displays all errors generated during analyzer communication. It helps laboratory staff identify and resolve integration problems quickly.
Dashboard KPIs
KPI | Description |
|---|---|
Total Errors | All errors in the system, acknowledged and unacknowledged. |
Unacknowledged | Errors not yet reviewed and acknowledged by a user. |
Critical | Errors at Critical severity requiring immediate attention. |
Last 24 Hours | Errors generated in the last 24-hour window. |
Error Table Columns
Column | Description |
|---|---|
Timestamp | Date and time the error was generated. |
Analyzer | Name of the analyzer that generated the error. |
Type | Error type (see Error Types below). |
Severity | Error severity level (see Severity Levels below). |
Message | Human-readable description of the error. |
Status | Whether the error has been acknowledged or is still unacknowledged. |
Actions | Per-row actions: Acknowledge, View Details. |
Filters
The Error Dashboard includes three filters above the table:
Filter | Options |
|---|---|
Error Type | All Types, Mapping, Validation, Timeout, Protocol, Connection |
Severity | All Severities, Critical, Error, Warning |
Analyzer | All (plus individual analyzer names for analyzers that have generated errors) |
Error Types
Error Type | Cause |
|---|---|
Connection | Unable to establish a network connection to the analyzer. |
Protocol | Message does not conform to the expected protocol format (ASTM or HL7). |
Mapping | A field in the message has no mapping, or the mapped OpenELIS test was not found. |
Validation | A result value failed validation rules (e.g., out-of-range numeric or unit mismatch with Reject enabled). |
Timeout | The connection attempt to the analyzer timed out. |
Severity Levels
Severity | Description |
|---|---|
Critical | Analyzer is not functioning. Patient results may be delayed or lost. |
Error | A specific message or result failed to process. Other messages may still be processing. |
Warning | A non-blocking issue that should be reviewed but does not stop message processing. |
Acknowledging Errors
Acknowledging an error marks it as reviewed. The analyzer's Error Pending status clears once all errors are acknowledged.
To acknowledge a single error, click Actions → Acknowledge on the error row.
To view full details before acknowledging, click Actions → View Details.
To clear all unacknowledged errors at once, click Acknowledge All at the top of the page (top-right, above the KPI cards).
4. Analyzer Types
Analyzer Types define the technical profile for a class of laboratory instrument, including the communication protocol, the plugin class that processes messages, and the regex pattern used to identify incoming messages. Analyzer Types must be configured before adding individual analyzer instances.
Analyzer Types Table Columns
Column | Description |
|---|---|
Name | Display name for the analyzer type (e.g., PENTRA 80, COBAS c311). |
Description | Free-text description of the analyzer model. |
Protocol | Communication protocol: ASTM, HL7, or FILE. |
Plugin Class | Fully qualified Java class name of the plugin that parses messages for this type. |
Identifier Pattern | Regex used to route incoming messages to this analyzer type. |
Generic Plugin | Whether this type uses the built-in generic plugin framework (Yes/No). |
Plugin Loaded | Whether the plugin class is currently loaded in the running system (Yes/No). |
Instances | Count of configured analyzer instances using this type. |
Status | Active or Inactive. Inactive types cannot be used when adding new analyzers. |
Creating a New Analyzer Type
Click Create New Analyzer Type (top-right of the Analyzer Types page).
Fill in the fields (see table below).
Click Create New Analyzer Type to save, or Cancel to discard.
Note: On some server configurations, the create action may display an error notification even when the type was successfully created. Refresh the page to confirm whether the new type appears in the list.
Field | Description |
|---|---|
Name | Short identifier for this type. Used in dropdowns when adding analyzer instances. |
Description | Optional description for documentation. |
Protocol | ASTM (default), HL7, or FILE. Choose based on how the physical analyzer communicates. |
Plugin Class Name | The fully qualified Java class name of the plugin JAR installed on the server. Not required when Generic Plugin is checked. |
Identifier Pattern | A Java regex to match incoming sample identifiers for this type (e.g., |
Generic Plugin | Check to use the built-in generic plugin framework. Plugin Class Name is not required when checked. |
Active | Defaults to checked. Uncheck to create the type in an inactive state (will not appear in analyzer instance dropdowns until activated). |
Notes & Cautions
Activating field mappings on a live (Active) analyzer applies only to new incoming messages. Previously received messages are not reprocessed.
An analyzer will not appear in the Active KPI count unless it has activated field mappings and a working network connection.
A Plugin Warnings KPI count appears on the Analyzers List when one or more analyzers have a missing plugin. Contact your system administrator to install the required plugin JARs.
The Plugin Type dropdown in the Add Analyzer form is populated from plugins installed on the server. If the list is empty, contact your system administrator.
Configure File Import is available for all analyzers regardless of protocol. Use it for any analyzer delivering results as flat files.
Deleting an analyzer permanently removes all its field mappings. Use Inactive status instead of deletion if you may need to restore the configuration later.
Analyzers in Offline status are automatically re-tested by the system. There is no manual "reconnect" button — the system will transition back to the appropriate status when the connection recovers.
Validated against OpenELIS Global v3.2.1.3 — March 2026
To Be Verified
The items below could not be directly observed in the OpenELIS Global v3.2.1.3 test instance because they require a live analyzer connection with real data flowing. Each item should be confirmed against a fully configured production or staging instance. Verification is done by a person navigating the live UI — no screenshots required.
A. Requires Live Analyzer Data
These features require actual analyzer messages to be received before the relevant UI elements become available.
Feature | Belongs In Section | What Is Missing | How to Verify |
|---|---|---|---|
QC Mapping sub-form |
| When a Control Test type field is selected in the left panel, the right panel may show a QC-specific mapping form linking the result to a QC record. The fields in this sub-form are undocumented. | With a live analyzer that sends Control Test fields, select one in the Field Mappings editor and document all right-panel form fields. |
Activation Confirmation Modal |
| The existing note mentions a warning and confirmation checkbox when activating mappings on an Active analyzer. The modal's full content — warning text, exact checkbox label, any secondary inputs — is not documented. | With an Active analyzer that has a saved draft mapping, click Save and Activate and record all modal contents including exact checkbox label. |
Melting Point field-specific configuration |
| No documentation on whether Melting Point fields have additional configuration beyond the standard mapping fields (temperature unit, curve parameters). | With a molecular analyzer sending Melting Point fields, select one in the Field Mappings editor and document all right-panel form fields. |
Custom field type behavior |
| The field type table lists Custom as "user-defined for non-standard outputs" but no details on what additional form fields appear or how routing works. | Select a Custom type field in the Field Mappings editor and record all form fields and available configuration options. |
Field mapping filter options |
| The filter dropdown labels are visible but the options list only populates when fields are present. Exact option labels (e.g., "Active Only", "Draft Only") need confirmation. | Open the field list filter dropdown when an analyzer with fields is loaded and record all options. |
B. Corrections Applied — Previously Marked Unverified
The following items from the previous "To Be Verified" list have been confirmed and corrected in the main documentation above:
Item | Result |
|---|---|
Error Types: Connection, Protocol, Mapping, Validation, Timeout | ✅ Confirmed. All five present. Order in dropdown: Mapping, Validation, Timeout, Protocol, Connection. |
Severity Levels: Critical, Error, Warning | ✅ Confirmed. All three present in Severity filter dropdown. |
Error Dashboard KPI cards: Total Errors, Unacknowledged, Critical, Last 24 Hours | ✅ Confirmed. All four cards present. |
Test Mapping options: "Detailed Parsing" and "Validate All" | ✅ Confirmed with corrections. Actual labels: "Show detailed parsing steps" and "Validate all mappings". Button is "Preview Mapping" (not "Submit"). |
"Save Test Case" option in Test Mapping | ❌ Not present. Removed from documentation. |
Protocol Version options: ASTM LIS2-A2, HL7 v2.3.1, HL7 v2.5 | ✅ Confirmed. All three options present. ASTM LIS2-A2 is default. |
"Acknowledge All" button — label and position | ✅ Confirmed. Button is at top-right of the Error Dashboard, above KPI cards. |
Analyzer Types list — Plugin Loaded and Instances columns | ✅ Confirmed. Plugin Loaded shows Yes/No. Instances shows numeric count. |
Analyzers List KPI cards: Total, Active, Inactive | ✅ Confirmed with correction. A fourth KPI — Plugin Warnings — also exists. |
Add Analyzer form — Load Analyzer Profile, Test Unit Assignment, Identifier Pattern fields | ❌ These fields do not exist in the Add Analyzer form. Removed from documentation. The Identifier Pattern field exists in the Analyzer Types form (Section 4), not here. |
Per-Analyzer Actions — Test Connection as row-level action | ✅ Confirmed. Test Connection appears in the overflow menu alongside Field Mappings, Configure File Import, Copy Mappings, Edit, and Delete. |
Configure File Import — only visible for FILE protocol analyzers | ❌ Incorrect. Configure File Import is available for all analyzers via the overflow menu. |
Qualitative code mapping — Default designation | Still unverified. Requires live analyzer with qualitative fields. |
Copy Mappings — blocked when source has no mappings | Still unverified. Requires two analyzers with mappings. |
To Be Verified section last updated: March 2026